KEW 6010B
MULTI-FUNCTION TESTER
INSTRUCTION MANUAL
KYORITSU ELECTRICAL INSTRUMENTS
WORKS,LTD.
1 Safety Warning ................................................................................. 1.
2 Instrument Layout ........................................................................... 4.
3 Features ......................................................................................... 5.
4 Specification ................................................................................... 7.
5 Continuity (resistance) tests ........................................................... 10 .
.5 1
Test procedure ..................................................................... 10
6.Insulation Tests ............................................................................. 12
.6 1
The nature of insulation resistance ....................................... 12
.6 1.2
Capacitive current ............................................................... 12
.6 1.3
Conduction current .............................................................. 13
.6 1.4
Surface leakage current ....................................................... 13
.6 1.5
Total leakage current ............................................................ 13
.6 2
Damage to voltage sensitive equipment ............................... 14
.6 3
Preparation for measurement ............................................... 15
.6 4
Insulation resistance measurement ...................................... 15
7 Loop Impedance Tests .................................................................. 17.
.7 1
Voltage measurement ........................................................... 17
.7 2
What is earth fault loop impedance? ..................................... 17
.7 3
Automatic over-temperature cut-out ..................................... 18
.7 4
The loop impedance test ...................................................... 19
.7 5
Loop impedance at 3 phase equipment ................................ 20
8.RCD/Uc Tests ............................................................................... 21
.8 1
Purpose of RCD Test ............................................................ 21
.8 2
What does the RCD test really do? ....................................... 21
.8 3
What is Uc? .......................................................................... 22
.8 4
Uc Testing ........................................................................... 23
.8 5
Operation of KEW 6010B RCD testing ................................. 23
.8 6
RCD testing ......................................................................... 23
.8 6.1
"NO TRIP" and "TRIP" test ................................................... 23
.8 6.2
"FAST TRIP" test .................................................................. 25
.8 6.3
Testing DC sensitive RCDs ................................................... 25
.8 6.4
Testing Auto Ramp ............................................................... 25
.8 7
Testing time delayed RCD's .................................................. 26
9 Store / Recall a measured result .................................................... 27.
.9 1
How to store the data ........................................................... 27
.9 2
Recall the stored data .......................................................... 28
.9 3
Delete the stored data .......................................................... 29
.9 4
Transfer the stored data to PC .............................................. 30
10 Battery / Fuse replacement ........................................................... 31.
11.General ........................................................................................ 32
12 Servicing ...................................................................................... 32.
13 Case, Strap and Shoulder-pad assembly ...................................... 33.
CONTENTS
—
1
—
1.
SAFETY WARNING
Electricity is dangerous and can cause injury and death. To avoid possible
electric shock, personal injury or damage of instrument, always treat it with
the greatest of respect and care. If you are not quite sure how to proceed,
stop and take advice from a qualified person.
1. This instrument must only be used by a competent and trained person and
operated in strict accordance with the instructions. KYORITSU will not
accept liability for any damage or injury caused by misuse or non-
compliance with the instructions or with the safety procedures.
2. It is essential to read and to understand the safety rules contained in the
instructions. They must always be observed when using the instrument.
3. This instrument is only intended for single phase operation at 230V AC
+10%, -15% phase to earth or phase to neutral operation, and then only for
Loop, RCD and Uc testing. For use in the continuity testing and insulation
testing modes this instrument must be used ONLY on circuits
which are de-energized.
4. Verify the tester's operation by measuring a known voltage before and
after using it.
5. When conducting tests do not touch any exposed metalwork associated
with the installation. Such metalwork may become live for the duration of
the test.
6. Never open the instrument case (except for fuse and battery
replacement and in this case disconnect all leads first) because dangerous
voltages are present. Only fully trained and competent electrical engineers
should open the case. If a fault develops, return the instrument to your
distributor for inspection and repair.
7. If the overheat symbol appears in the display " " disconnect the
instrument from the mains supply and allow to cool down.
8. For loop impedance tests to prevent unwanted tripping during loop testing
all residual current devices(RCDs) must be taken out of the circuit and
temporarily replaced with a suitably rated MCB unit. The RCD must be
replaced after the loop test is completed.
9. If abnormal conditions of any sort are noted (such as a faulty display,
unexpected readings, broken case, cracked test leads, etc) do not use the
tester and return it to your distributor for repair.
10. For safety reasons only use accessories (test leads, probes, fuses, cases,
etc) designed to be used with this instrument and recommended by
KYORITSU. The use of other accessories is prohibited as they are unlikely to
have the correct safety features.
11.
When testing, always be sure to keep your fingers behind the protective
fingerguard on the test leads.
12.
During testing it is possible that there may be a momentary degradation of
the reading due to the presence of excessive transients or discharges on
the electrical system under test. Should this be observed, the test must be
repeated to obtain a correct reading. If in doubt, contact your distributor.
13.
The sliding shutter on the back of the instrument is a safety device. The
instrument should not be used if it is damaged or impaired in any way, but
Termékspecifikációk
Márka: | Kyoritsu |
Kategória: | multiméter |
Modell: | KEW 6010B |
Szüksége van segítségre?
Ha segítségre van szüksége Kyoritsu KEW 6010B, tegyen fel kérdést alább, és más felhasználók válaszolnak Önnek
Útmutatók multiméter Kyoritsu
14 Január 2025
15 December 2024
15 December 2024
15 December 2024
27 Szeptember 2024
20 Szeptember 2024
27 Augusztus 2024
27 Augusztus 2024
27 Augusztus 2024
27 Augusztus 2024
Útmutatók multiméter
- multiméter PeakTech
- multiméter Milwaukee
- multiméter Maxwell
- multiméter Emos
- multiméter Voltcraft
- multiméter Vimar
- multiméter Digitus
- multiméter GW Instek
- multiméter Owon
- multiméter Aim TTi
- multiméter Gossen Metrawatt
- multiméter Amiko
- multiméter Klein Tools
- multiméter Chauvin Arnoux
- multiméter Workzone
- multiméter Beha-Amprobe
- multiméter Qian
- multiméter Joy-It
- multiméter Extech
- multiméter Testo
- multiméter Powerfix
- multiméter Biltema
- multiméter REV
- multiméter BENNING
- multiméter Alecto
- multiméter Yato
- multiméter Velleman
- multiméter TFA
- multiméter Monacor
- multiméter Elro
- multiméter Fluke
- multiméter Uni-T
- multiméter Metrix
- multiméter Abus
- multiméter Megger
- multiméter Testboy
- multiméter Sonel
- multiméter Wiha
- multiméter Schneider
- multiméter Brandson
- multiméter Silverline
- multiméter Rigol
- multiméter Projecta
- multiméter Multimetrix
- multiméter Weidmüller
- multiméter Laserliner
- multiméter Appa
- multiméter Amprobe
- multiméter Brennenstuhl
- multiméter Ferm
- multiméter Tacklife
- multiméter Somogyi
- multiméter Profile
- multiméter Steren
- multiméter Flir
- multiméter Hager
- multiméter Clas Ohlson
- multiméter Etekcity
- multiméter Greenlee
- multiméter Högert
- multiméter Bruder Mannesmann
- multiméter Keithley
- multiméter Metrel
- multiméter Digi-tool
- multiméter Skandia
- multiméter Elma
- multiméter Hubinont
- multiméter Kewtech
- multiméter Strex
- multiméter Mastech
- multiméter MBS
- multiméter MGL Avionics
- multiméter PCE Instruments
- multiméter Micronta
- multiméter CEM
- multiméter Horex
- multiméter Cosinus
- multiméter Kurth Electronic
- multiméter Testec
- multiméter IWH
- multiméter Weltron
- multiméter Sanwa
- multiméter Rohde & Schwarz
- multiméter EEVBlog
- multiméter Kingcraft
- multiméter HT Instruments
- multiméter Caltek
- multiméter Elworks
- multiméter Bearware
- multiméter Noyafa
Legújabb útmutatók multiméter
13 Január 2025
10 Január 2025
3 Január 2025
1 Január 2025
29 December 2024
29 December 2024
29 December 2024
29 December 2024
23 December 2024
22 December 2024