INSTRUCTION MANUAL
DIGITAL CLAMP METER
KEW 2055
1.Features
●
Designedtomeetinternationalsafetystandards.
IEC61010-1.IEC61010-031:2002&IEC61010-2-032
MeasurementCategory(CAT)IV600V
PollutionDegree2
●
Double molded main body provides comfortable
singlehandedgrip
●
DataHoldFunction
●
LCD Backlight function to facilitate working at
dimlylitsituations.
●
RELfunctiontoindicatemeasurementvariation.
(Current,voltage,Resistancemeasurement)
●
MIN/MAX function enables easy reading of min &
maxvalueduringmeasurement.
●
WithContinuity&DiodeCheckFunction
●
NCV
(
Non Contact Voltage
)
Function for wiring
check.
●
600Vinputprotection.
●
Sleepfunctiontoextendbatterylife.
●
WithBarGraph,6039counts
2. Safety Warnings
This instrument has been designed, manufactured and
testedaccordingtoIEC61010:Safetyrequirementsfor
Electronic Measuring apparatus, and delivered in
the best condition after passed the inspection. This
instruction manual contains warnings and safety
rules which must be observed by the user to ensure
safeoperation of the instrument and retain it in safe
condition.
Therefore, read through these operating instructions
beforeusingtheinstrument.
WARNING
●
Read through and understand the instructions
contained in this manual before using the
instrument.
●
Keep the manual at hand to enable quick
referencewhenevernecessary.
●
Theinstrumentistobeusedonlyinitsintended
applications.
●
U n d e r s t a n d a n d f o l l o w a l l t h e s a f e t y
instructionscontainedinthemanual.
Failure tofollow the instructions may cause injury,
instrument damage and/or damage to equipment
under test. Kyoritsu is by no means liable for
any damage resulting from the instrument in
contradictiontothiscautionarynote.
Thesymbol indicatedontheinstrumentmeansthat
theusermustrefertotherelatedpartsinthemanual
forsafe operation ofthe instrument. It isessential to
read the instructions wherever the symbol appears
inthemanual.
DANGER is reserved for conditions and actions
that are likely to cause serious or fatal injury.
WARNING is reserved for conditions and actions
that can cause serious or fatal injury.
CAUTION is reserved for conditions and actions
that can cause injury or instrument damage.
●
Marks listed in the table below are used on this
instrument.
Usermustrefertothemanual.
Instrumentwithdoubleorreinforcedinsulation
Indicates that this instrument can clamp on
bare conductors when measuring a voltage
corresponding to the applicable measurement
category,whichismarkednexttothissymbol.
AC
DC
AC&DC
Thisinstrumentsatisfiesthemarkingrequirement
defined in the WEEE Directive (2002/96/EC).
This symbol indicates separate collection for
electricalandelectronicequipment.
DANGER
●
Never make measurement on a circuit in which
voltageoverAC600Vexists.
●
Do not attempt to make measurement in the
presenceof flammable gasses. Otherwise, the use
of the instrument may cause sparking, which can
leadtoanexplosion.
●
Transformer jaw tips are designed not to short
the circuit under test. If equipment under test
has exposed conductive parts, however, extra
precaution should be taken to minimize the
possibilityofshorting.
●
Neverattempttousetheinstrumentifitssurface
oryourhandiswet.
●
Do not exceed the maximum allowable input of
anymeasuringrange.
●
N e v e r o p e n t h e B a t t e r y c o v e r d u r i n g a
measurement.
●
The instrument is to be used only in its intended
applications or conditions. Otherwise, safety
functionsequippedwiththeinstrumentdoesn
twork,
and instrument damage or serious personal injury
maybecaused.
●
Verify proper operation ona known source before
use or taking action as a result indication of the
instrument.
●
Keepyourfingersandhandsbehindtheprotective
fingerguardduringmeasurement.
WARNING
●
Never attempt to make measurement if any
abnormal conditions, such as broken case and
exposedmetalpartsarefoundontheinstrument.
●
Do not rotate the Function Switch while the test
leadsarebeingconnected.
●
Do not install substitute parts or make any
modification to the instrument. For repair or
re-calibration, return the instrument to your local
distributorfromwhereitwaspurchased.
●
Donot try to replace the batteries if the surface
oftheinstrumentiswet.
●
Disconnect all the cords and cables from the
object under test and power offthe instrument
before opening the Battery Cover for Battery
replacement.
●
Verify proper operation ona known source before
useortakingactionasaresultoftheindicationof
theinstrument.
●
Use appropriate personal protective equipment
such as insulating gloves, Insulating boots, and
safetyglasses.
●
Stop using the test lead if the outer jacket is
damaged and the inner metal or color jacket is
exposed.
CAUTION
●
SettheFunctionSwitchtoanappropriateposition
beforestartingmeasurement.
●
Firmlyinsertthetestleads.
●
Disconnectthetestleadsfromtheinstrumentfor
currentmeasurement.
●
Do not exposethe instrument to the direct sun,
hightemperatureandhumidityordewfall.
●
Altitude 2000m or less. Appropriate operating
temperatureiswithin0
℃
and40
℃
.
●
This instrument isn
t dust & water proofed. Keep
awayfromdustandwater.
●
Be sure to power off the instrument after use.
Whentheinstrumentwill notbein usefora long
period, place it in storage after removing the
batteries.
●
Use a cloth dipped in water or neutraldetergent
for cleaning the instrument. Do not use abrasives
orsolvents.
Measurement Category
To ensure safe operation of measuring instruments,
IEC 61010 establishes safety standards for various
electrical environments, categorized as O to CAT IV,
and called measurement categories. Higher-numbered
categories correspond to electrical environments with
greatermomentary energy,so a measuringinstrument
designedforCAT III environmentscanendure greater
momentaryenergythanonedesignedforCATII.
O
:Circuits which are not directly connected
tothemainspowersupply.
CAT II
:Electrical circuits of equipment connected
toanACelectricaloutletbyapowercord.
CAT III
:Primary electrical circuits of the
equipment connected directly to the
distribution panel, and feeders from the
distributionpaneltooutlets.
CAT IV
:The circuit from the service drop to the
service entrance, and to the power meter
andprimaryover-currentprotectiondevice
(distributionpanel).
3. Specification
3-1. Measuring range & accuracy
(accuracyguaranteedat23
℃ ±
5
℃
.humidity45
〜
85%)
ACCurrent600A,1000AFunction
Function Measuring
Range Accuracy
600A 0-600.0A
±
1.5%rdg
±
5dgt
(50/60Hz)
±
3.5%rdg
±
8dgt
(40-400Hz)
1000A 0-1000A
DCCurrent600A,1000AFunction
Function Measuring
Range Accuracy
600A
0-600.0A
±
1.5%rdg
±
5dgt
1000A
0-1000A
ACVoltageFunction
(
Auto-ranging,Inputimpedance:approx.10M
Ω)
Range Measuring
Range Accuracy
6/60/600V 1-600.0V
±
1.3%rdg
±
4dgt(50/60Hz)
±
3.0%rdg
±
5dgt(40
〜
400Hz)
DCVoltageFunction
(
Auto-ranging,Inputimpedance:approx.10M
Ω)
Range Measuring
Range Accuracy
600mV/6/
60/600V 0-600.0V
±
1.0%rdg
±
3dgt
Resistance
(
Continuity/DiodeCheck
)
Function
Range Measuring
Range Accuracy
600
Ω
/6k/
60k/600k
Ω
6M
Ω
0-6M
Ω ±
1.0%rdg
±
5dgt
60M
Ω
6.00M-
60.00M
Ω±
5%rdg
±
8dgt
ContBuzzer 0-600.0
Ω
Thresholdvalue:60
Ω±
30
Ω
Diode Testvoltage:0-2V
Frequency/ DUTY Function
(
Auto-ranging f or
Frequency
)
Range Measuring
Range Accuracy
ACA 40Hz
−
400Hz
±
0.5%rdg
±
5dgt
ACV 1Hz
〜
10kHz
0.1-99.9%
(Pulsewidth/Pulseperiod)
±
2.5%rdg
±
5dgt
Note:Measurableinputsare:40Vrms@ACVor
60Arms@AC600A,350A@AC1000ARange
3-2. General Specification
●
Modeofoperation
:
Δ
∑mode
●
Display
:max.6039counts
(Frequency:9999)&Bargraph
●
Over-range indication : "OL" displayed when exceeding
the measuring range. (except for AC/DCV and 1000A
Function)
●
Rangeswitching :
Auto-ranging/Voltage,ResistanceRange
Singlerange/Continuity,Diodecheckand
DUTY
●
Samplerate
:threetimespersecond
●
Functionalconstruction:
OFF/ACA/DCA/ACV/DCV/
Ω
●
Keys :
SELECT(AC/DCswitching&/
Ω
/ / ),REL
Δ
,
Hz/DUTY,MIN/MAX,HOLD/BackLight
●
Powersource :DC3V/R03(UM-4)x2pcs
●
Lowbatterywarning
:
" "markisdisplayedat
2.4V
±
0.15Vorless.
●
Temperature & humidity : 23
℃ ±
5
℃
, relative humidity
accuracyguaranteed85%orless(nocondensation)
●
Operatingtemperature:0
〜
40
℃
,relativehumidity
&humidityrange85%orless(nocondensation)
●
Storagetemperature :-20
〜
60
℃
,relativehumidity
&humidityrange85%orless(nocondensation)
●
Currentconsumption:approx.12mA
●
SleepFunction:Automaticallypoweredoffinabout
15minafterthelastFunctionswitchoperation.
Press any key or rotate the Function Switch from
OFFtoanypositiontoexitfromtheSleepstate.
●
Locationforuse
Outdooruse,Altitudeupto2000m
●
ApplicableStandards
IEC61010-1,61010-2-032,61010-2-033
MeasurementCATIV600VPollutiondegree2
IEC61010-031
EMC
:
EN61326-1
・
EN55022
・
EN61000-4-2(performancecriterionB)
・
EN61000-4-3(performancecriterionB)
●
RoHS
:
EN50581
●
OverloadProtection
CurrentRange :1200AAC/DC/10sec
VoltageRange :720VAC/DC/10sec
ResistanceRange :600VAC/DC/10sec
●
WithstandVoltage
6720VAC
(
TRMS50/60Hz
)
/5sec
(betweenJawsandelectricalcircuit/betweeninternal
circuitandenclosure)
●
InsulationResistance:10M
Ω
ormore/1000V
(
betweenelectricalcircuitandenclosure
)
●
Conductorsize
approx.40mm
●
Dimension
approx.254(L)
×
82(W)
×
36(D)mm
●
Weight:approx.:310g
●
Accessories
TestLeads Model7066A/1set
BatteryR03
(
UM-4
)
/2pcs
InstructionmanualEnglish,Japanese/1pce
CarriyngCaseModel9094/1pcs
3-3. Function Keys
The"
●
"markshowsavailablefunctionateachRange.
DCCurrentMeasurement
HOLD SELECT ZERO Hz/
DUTY MAX/
MIN
ACA
●●●●●
ACV
●
-
●●●
DCA
●●●
-
●
DCV
●
-
●
-
●
Ω●●●
-
●
-
●
---
-
●
---
4. Preparation for measurement
4-1. Checking Battery Voltage
Set the Function Switch to any position other than
"
OFF
".When thedisplayis clearwithout "
BATT
" mark,
showing, battery voltage is enough. When the display
is blank or
"
BATT
" mark is indicated, replace the
batteriesaccordingtoSection7,BatteryReplacement.
CAUTION
The Sleep feature automatically powers the
instrumentoffinabout 15minafterthe lastswitch
or key operation. Therefore, the display may be
blankevenwiththeFunctionSwitchsettoaposition
other than "
OFF
". To operate the instrument in this
case, turn the switch back to the "
OFF
" position,
thentoanyotherposition,orpressanykey.Replace
the batteries if nothing was displayed after above
operations.
4-2. Checking Switch Setting & Operation
Confirm the Function Switch is set to the correct
position, the instrument is set to the correct
measurementmode,
and the Data hold function is disabled. Otherwise,
desiredmeasurementcannotbemade.
5. Measurement
5-1. AC Current Measurement
DANGER
●
Never make measurement on a circuit in which
voltage over AC600V exists to avoid getting
electricalshock.
●
Transformer jaw tips are designed not to short
the circuit under test. If equipment under test has
exposedconductiveparts,however,extraprecaution
should be taken to minimize the possibility of
shorting.
●
Do not make measurement with the Battery Cover
removed.
●
Disconnect the test leadsfrom the instrument for
currentmeasurement.
●
Keep yourfingersand handsbehindthe barrieron
theinstrumentduringameasurement.
(1)Set the Function Switch to "
600A
" or "
1000A
"
position.
AChasbeenselectedbydefault;presstheSELECT
key, when DC has been selected, to change it to
AC.AC mark isdisplayed at theupper left onthe
display.
(2)Pressthetriggertoopenthetransformerjawsand
clampthemontotheoneconductorundertest,then
take the reading on the display. Pressing the "
Hz/
DUTY
" Key switches the indication in following
sequence.
AC Current
⇨
Hz
⇨
DUTY
Hz/DUTY
Function requires 60A or more at AC600A
Rangeand350AormoreatAC1000Arange.
CAUTION
●
Max conductor size for KEW2055 is approx
dia. 40mm. During current measurement, keep
the trans former jaws fully closed. Other wise,
accuratemeasurementscannotbetaken.
5-2. DC Current Measurement
DANGER
●
Never make measurement on a circuit in which
voltage over DC600V exists to avoid getting
electricalshock.
●
DonotmakemeasurementwiththeBatteryCover
removed.
●
Keepyourfingersandhandsbehindthebarrieron
theinstrumentduringameasurement.
(1)Set the Function Switch to "
600A
" or "
1000A
"
position. AC has been selected by default; press
the SELECT key, when AC has been selected, to
changeittoDC.
DC mark is displayed at the upper left on the
display.
Barrier
O: Device which is
not directly
connected to the
mains power supply
Protectivefingerguard(Barrier)
:
It is a part providing protection against electrical
shock and ensuring the minimum required air and
creepagedistances.
Cap
:
UncappedconditionforCATIIenvironment
CappedconditionforCATIII/IVenvironments
TheCapshuldbefirmlyattachedtotheprobes.
Whenthe instrument and the testlead are combined
andusedtogether,whicheverlowercategoryeitherof
thembelongstowillbeapplied.
Cap
Protective fingerguard

5-4. DC Voltage Measurement
DANGER
●
Never make measurement on a circuit in which
voltage over DC600V exists to avoid getting
electricalshock.
●
Do notmakemeasurementwith theBatteryCover
removed.
●
Keepyourfingersbehindtheprotectivefigerguard
ontheinstrumentduringmeasurement.
(1)SettheFunctionSwitchto"
DCV
"position.
(2)ConnecttheredtestleadtoV/
Ω
terminalandthe
blacktestleadtoCOMterminal.
(3)Connect the red and black test leads to the
positive (+)and negative (-) sides of the circuit
under test respectively. Take the reading on the
display. If the connection is reversed, the display
indicatesthe"-"mark.
5-5. Resistance/ Cont/ Diode Measurement
DANGER
●
Neverusetheinstrumentonanenergizedcircuit.
●
Do not make measurement with the Battery
Coverremoved.
Resistance
(1)Set the Function Switch to "
Ω/ Cont/ Diode
"
position.
(2)ConnecttheredtestleadtoV/
Ω
terminalandthe
black test lead to COM terminal. Confirm "OL" is
indicatedon thedisplay,and thenshort-circuitthe
tipsoftestleadstomaketheindicationzero.
(3)Connect the test leads to the both ends of the
resistorundertest.
(4)Takethereadingonthedisplay.
CAUTION
●
Evenif shortthe testlead tips,indicatedvalue
may not be zero. But this is because of the
resistanceoftestleadsandnotafailure.
●
Whentestleadsareopen,"OL"isindicatedon
thedisplay.
●
Keep your fingersand hands behindthe barrier
duringmeasurement.
Continuity
(1)Set the Function Switch to "
Ω/ Cont/ Diode
"
position. "
Ω
" has beenselected by default; press
theSELECTkeytochangeitto"
Continuity
".
Resistance Cont Diode⇨ ⇨
(2)ConnecttheredtestleadtoV/
Ω
terminalandthe
black test lead to COM terminal. Confirm "
OL
" is
indicated on the display and short-circuit the tips
of test leads. Indication should become zero and
buzzersounds.
(3)Connect the test leads to the both ends of the
conductor under test. The buzzer sounds, if the
resistanceundertestis100
Ω
orless.
Diode
(1)Set the Function Switch to "
Ω/ Cont/ Diode
"
position. "
Ω
" has been selected by default; press
theSELECTkeytochangeitto"
Diode
"
Resistance Cont Diode⇨ ⇨
(2)ConnecttheredtestleadtoV/
Ω
terminalandthe
blacktestleadtoCOMterminal.
(3)ConnecttheredandblacktestleadstotheAnode
and Cathode of the diode under test respectively.
Take the readingon the display. If theconnection
isreversed,thedisplayindicates"
OL
".
CAUTION
●
Some ofdiodes cannotbe tested.Indication on
thedisplay willbe"OL".
(Zenerdiode, LEDand
soon)
6. Other functions
6-1. Sleep Function
(1)This is a function to prevent the instrument from
beingleftpoweredoninordertoconservebattery
life.
This function causes the instrument to enter
Sleep mode about 15 minutes after the last
key operation. To exit the Sleep mode, turn the
Functionswitchto"
OFF
",thentoanyotherposition,
orpressanyKey.
(2)SleepFunctionisdisabledwhen;MIN/MAX
Function is selected. Continuous measurement is
made with the Sleep Function being disabled. To
activate Sleep Function again, disable the MIN/
MAXFunction.
CAUTION
●
The instrument consumes small amount of
battery power in the Sleep mode. Set the
FunctionSwitchtotheOFFpositionafteruse.
6-2. HOLD Key
(1)DataHoldFunction
This is a function to freeze the measured value
on the display. Pressthe "HOLD" key to freeze the
reading.
The reading will be held regardless of subsequent
variationininput."
H
"isindicatedontheupperleft
cornerofthe displaywhilethe instrumentis inthe
DataHoldmode.ToexitDataHoldmode,pressthe
"HOLD"keyagain.
CAUTION
●
Held readings are released when Sleep
Functionisactivatedwhiletheinstrumentisin
theDataHoldmode.
(2)BacklightON/OFF
PressingtheHOLDkey2secormorelightsupthe
Backlight. Pressing the HOLD key 2 sec or more
againturnsofftheBacklight.
6-3. NCV Function
Red LED on the upper area on the Panel lights up
at all functions except for OFF when electric field
exceeding 100V is detected by the sensor installed in
theJaws.
It indic ates a pre se nce o f
voltageinanelectricalcircuit
or equipment without touching
them.
N C V S e n s o r c a n d e t e c t
electrical field only from the
direction indicated in theright
figure.
Put the fixed element (left
side) closer to the conductor
undertest.
Detection agai nst i n-wall
outletisimpossible.
DANGER
●
The LED may not light up due to installation
conditionofelectricalcircuitorequipment.Never
touch the circuit under test to avoid possible
dangereveniftheLEDforNCVdoesn
tlightup.
●
Check the functionality of LED on a well-known
powersupplypriortomeasurement.WhentheLED
doesn
tlightup,donotmakemeasurement.
●
NCV indication is affected by external voltage,
howtoholdorplacetheinstrument.
●
Keep your fingers and hands behind the barrier
duringmeasurement.
6-4. MIN/MAX Function
CAUTION
●
SELECT, ZERO, Hz/DUTY keys are disabled
whileMIN/MAXFunctionisbeingactivated.
(1)
AC/DC Current Range
Pressing the MIN/MAX Key at 600A & 1000A
Function enables min or max value measurement.
Pressthe MIN/MAXKey toselectMAX orMIN.The
max ormin valuewithin measuring rangeis being
held until this function is disabled. "MIN" or "MAX"
is indicated on the display while this function is
being activated. To disable this function, press
down the MIN/MAX Key at least 2 sec or change
functions.
(2)
AC/DC Voltage Range
CAUTION
●
Pressing the MIN/MAX Key without applying
voltage disables the Auto-ranging function and
fixes the Range to 6V. Connect the test leads to
the circuit under test andpress the MIN/MAX Key
after an appropriate range is selected by Auto-
rangingfunction.
Pressing the MIN/MAX Key enables min or max value
measurement. Press the MIN/MAX Key to select MAX
orMIN.Themaxorminvaluewithinmeasuringrange
is being held until this function is disabled. "MIN" or
"MAX"isindicatedonthedisplaywhilethisfunctionis
beingactivated.
Todisablethisfunction,pressdowntheMIN/MAXKey
atleast2secorchangefunctions.
6-5. ZERO Function
CAUTION
●
MIN/MAX, PEAK keys are disabled while ZERO
Functionisbeingactivated.
Zero AdjustmentFunction at CurrentRange "
Δ
" mark
istobeindicatedattheupperrightonthedisplaywhile
ZEROfunctionisbeingoperated.Indicationofrelative
value at Current, Voltage, Resistance: Pressing the
ZERO Key indicates REL (relative value) Press the
ZERO Key to save the initial value at the start of
measurementasareferencevalue.Thenthedifference
between the later measured valuesand the reference
valueisindicatedonthedisplay.
The Auto-ranging function is disabled, while this
function isbeing activated,and the Rangeis fixed to
theRangeselectedatthestartofmeasurement.
Relativevalueisindicatedwithinfollowingranges.
(
Measuringrange
)
=(Full-scalevalueatthefixedRange)‒(Initialvalue)
Todisablethis function,press downtheMIN/MAXKey
atleast2secorchangefunctions.
6-6. Over-flow indication
When the input exceeds the measuring range at each
FunctionotherthanVoltageand1000ARange,"OL"or
"-OL"isindicatedonthedisplay.
7. Battery Replacement
WARNING
●
MIN/MAX, PEAKkeys aredisabled while ZERO
Functionisbeingactivated.
CAUTION
●
Donotmixoldandnewbatteries.
●
Installbatteriesincorrectpolarityasindicated
intheBatteryCompartment.
Replace the batteries when a Low Battery
Voltagewarning "BATT" mark is indicatedon the
display. Note that when the battery is completely
exhausted,thedisplayblankswithout"
BATT
"mark
shown.
(1)SettheFunctionSwitchto"OFF"position.
(2)Unscrew and remove the Battery Compartment
Coveronthebottomoftheinstrument.
(3)Replace the batteries observing correct polarity.
UsenewR03(AAA)orLR03/1.5Vbatteries.
(4)Install the Battery Compartment and tighten the
screws.
(2)With the transformer jaws closed and without
clamping them onto the conductor, press the
"
ZERO
" key to zero adjust the display. (mark is
displayedattheupperrightonthedisplay.)
(3)Pressthetriggertoopenthetransformerjawsand
clampthem ontothe oneconductor undertest, the
conductorshouldbeatthecenterofthejaws,then
takethereadingonthedisplay.
(4)Set theFunction Switch to anappropriate position
accordingtocurrentundertest.
(5)Pressing the "
ZERO
" key again releases "
ZERO
"
function. (mark at the upper right on the display
disappears.)
CAUTION
●
When the current flows from the upside (the
displayside)totheundersideoftheinstrument,the
polarityofthereadingispositiveandviceversa.
5-3. AC Voltage Measurement
DANGER
●
Never make measurement on a circuit in which
voltage over AC600V exists to avoid getting
electricalshock.
●
Do notmake measurementwiththe BatteryCover
removed.
●
K e e p yo u r f i ng e r s b e h i n d t h e p r o t e c t i ve
fingerguardontheinstrumentduringmeasurement.
(1)SettheFunctionSwitchto"
ACV
"position.
(2)ConnecttheredtestleadtoV/
Ω
terminalandthe
blacktestleadtoCOMterminal.
(3)Connect the test leads to the circuit under test.
Take the reading on the display. Pressing the
"
Hz/DUTY
" key while reading is indicated on
the display switches the indication in following
sequence.
AC Voltage Hz DUTY⇨ ⇨
CAUTION
●
Hz/DUTYFunctionrequiresAC40Vorhigher.
●
To measure a frequency, measure the voltage on
theelectricalcircuitinadvance.
●
Then press the Hz/DUTY key to enter into
frequencymeasurement.
●
Readings of frequency may fluctuate or be
influencedundernoisyenvironment.
Anode
Cathode
92-1801D12-18
8. Maintenance
●
Cleaning
Use a cloth dipped in water or neutral detergent for
cleaningtheinstrument.
Donotuseabrasivesorsolvents.Otherwise,instrument
getdamaged,deformedordiscolored.
Termékspecifikációk
Márka: | Kyoritsu |
Kategória: | multiméter |
Modell: | KEW 2055 |
Szüksége van segítségre?
Ha segítségre van szüksége Kyoritsu KEW 2055, tegyen fel kérdést alább, és más felhasználók válaszolnak Önnek
Útmutatók multiméter Kyoritsu

14 Január 2025

15 December 2024

15 December 2024

15 December 2024

27 Szeptember 2024

20 Szeptember 2024

27 Augusztus 2024

27 Augusztus 2024

27 Augusztus 2024

27 Augusztus 2024
Útmutatók multiméter
- multiméter PeakTech
- multiméter Milwaukee
- multiméter Parkside
- multiméter Maxwell
- multiméter Emos
- multiméter Voltcraft
- multiméter Vimar
- multiméter Digitus
- multiméter GW Instek
- multiméter Owon
- multiméter Aim TTi
- multiméter Gossen Metrawatt
- multiméter Amiko
- multiméter Klein Tools
- multiméter Chauvin Arnoux
- multiméter Workzone
- multiméter Beha-Amprobe
- multiméter Qian
- multiméter Joy-It
- multiméter Extech
- multiméter Testo
- multiméter Powerfix
- multiméter Biltema
- multiméter BaseTech
- multiméter REV
- multiméter BENNING
- multiméter Alecto
- multiméter Yato
- multiméter Velleman
- multiméter TFA
- multiméter Monacor
- multiméter Elro
- multiméter Fluke
- multiméter Uni-T
- multiméter Metrix
- multiméter Abus
- multiméter Megger
- multiméter Testboy
- multiméter Sonel
- multiméter Wiha
- multiméter Schneider
- multiméter Brandson
- multiméter Stanley
- multiméter Silverline
- multiméter Rigol
- multiméter Projecta
- multiméter Multimetrix
- multiméter Weidmüller
- multiméter Laserliner
- multiméter Appa
- multiméter Amprobe
- multiméter Brennenstuhl
- multiméter Ferm
- multiméter Tacklife
- multiméter Somogyi
- multiméter Profile
- multiméter Steren
- multiméter Perel
- multiméter Flir
- multiméter Proline
- multiméter Hager
- multiméter Topcraft
- multiméter Clas Ohlson
- multiméter Jung
- multiméter Etekcity
- multiméter Cablexpert
- multiméter Greenlee
- multiméter Högert
- multiméter Wago
- multiméter Bruder Mannesmann
- multiméter Keithley
- multiméter Metrel
- multiméter Digi-tool
- multiméter Skandia
- multiméter Elma
- multiméter Hubinont
- multiméter Kewtech
- multiméter Strex
- multiméter Mastech
- multiméter MBS
- multiméter MGL Avionics
- multiméter PCE Instruments
- multiméter Micronta
- multiméter CEM
- multiméter Horex
- multiméter Cosinus
- multiméter Kurth Electronic
- multiméter Testec
- multiméter IWH
- multiméter Weltron
- multiméter Sanwa
- multiméter Rohde & Schwarz
- multiméter EEVBlog
- multiméter Kingcraft
- multiméter HT Instruments
- multiméter Caltek
- multiméter Elworks
- multiméter Bearware
- multiméter Noyafa
Legújabb útmutatók multiméter

3 Április 2025

3 Április 2025

3 Április 2025

3 Április 2025

3 Április 2025

3 Április 2025

3 Április 2025

28 Március 2025

26 Március 2025

13 Január 2025