INSTRUCTION MANUAL
UPPER ARM SEMI-AUTOMATIC
BLOOD PRESSURE MONITOR
Model: 2004
22-32-cm
2~3cm
TYPE:ACMNP-1
Blood Pressure Monitor
START
MEMORY
-1-
95
mm Hg
Hypertension
Borderline
Normal
140 160 mm Hg
90
Diastolic
WHO Blood Pressure Classification
MAIN UNIT
INTRODUCTION
Thank you for purchasing the upper arm Semi-Automatic Blood Pressure Monitor. This
remarkable instrument is ideal for people who frequently monitor their own blood pressure
because it's easy to use.
Please read this instruction manual thoroughly before using the M
onitor. For specific
information on your own blood pressure, CONTACT YOUR DOCTOR.
The World Health Organisation (WHO) developed the following Blood Pressure
Classification. This classification, however, is only a general guideline because blood
pressure varies from person to person according to age, weight, and health status.
CONSULT YOUR DOCTOR TO DETERMINE YOUR NORMAL BLOOD PRESSURE.
Note: There is not an universally accepted definition of hypotension (low blood pressure),
but a systolic pressure below 99 mmHg is usually regarded as hypotension.
FEATURES OF THE PRODUCT
1. Using the fuzzy-logic measurement.
2. Memory can store 128 measurements.
3. Easy to use buttons.
4. Large and clear display.
5. Upper arm cuff is easy to wrap correctly.
6. Desktop style for easy measurement.
7. Automatically turns off (within 2.5 minutes) to save power.
1. Systolic Value
2. Diastolic Value
3. Pulse/min Value
4. MEMORY Button
5. ON/OFF/START Button
6. Blood pressure Display
7. Air Tube
8. Upper Arm Cuff
9. Inflation Bulb
10. Exhaust Valve
11. Air Tube Hole
Systolic
Normal
Borderline
Hypertension
Systolic (mmHg)
less than 139
140 to 159
more than 160
Diastolic (mmHg)
less than 89
90 to 94
more than 95
-2-
9
8
7
10
11
6
5
4
3
2
1
22-32-cm
2~3cm
TYPE:ACMNP-1
Blood Pressure Monitor
SYMBOLS ON DISPLAY
SYS
DIA
PULSE
-min
mmHg
Kpa
SYSTOLIC
MEMORY
DIASTOLIC
INFLATION
DEFLATION
LOW BATTERY
HEART BEAT PULSE UNIT
ATTENTION OF ERROR
VALUE OF SYSTOLIC
VALUE OF DIASTOLIC
UNIT OF PRESSURE
VALUE OF PULSE PER MINUTE
START
MEMORY
-3-
1. Push the cuff over the left upper arm so that
the tubes point in the direction of the lower
arm and palm.
2. Lay the cuff on the arm as illustrated.
Make certain that the lower edge of the cuff
lies approximately 2 to 3 cm (3/4~1") above
the elbow and that the rubber tube leaves
the cuff on the inner side of the arm.
Important! The mark (about 3 cm long bar)
must lie exactly over the artery which runs
down the inner side of the arm.
3. Tighten the free end of the cuff and close
the cuff with the closer.
4. There must be no free space between
the arm and the cuff as this would influence
the result. Clothing must not restrict the arm.
Any piece of clothing which does (e.g. a
pullover) must be taken off.
5. Secure the cuff with the Velcro closer in
such a way that it lies comfortably and is not
too tight. Lay the arm on the table (palm
upwards) so that the cuff is at the same
height as the heart. Make sure that the tube
is not kinked.
6. Remain seated quietly for two minutes
before you begin the measurement.
HOW TO USE THE UPPER ARM CUFF
2~3cm (3/4~1")
-4-
INSTALL BATTERIES
During the battery symbol appears on the display or nothing is display while inflation during
measurement, you should replace with new batteries.
1. Slide the battery cover in the direction of the arrow to open.
2. Insert four "AA" batteries in the correct polarity direction.
3. Replace the battery cover.
4. With the use of four new "AA" batteries, you can measure approximately 250 times
at room temperature of 22OC and inflating to around 170mmHg once a day. But the
included batteries only for test purpose may not last as specified long period.
batteries, please do not mix the new and old battery together.
Four "AA" batteries
Disposal of empty battery to the authorized
collecting party subject to the regulation of
each individual territory.
5. When the battery low icon displays on the LCD, please replace four new
"AA"
6. If the unit will not be used for a long period, please take out all batteries to avoid liquid
leakage damaging the unit.

Termékspecifikációk

Márka: Bioland
Kategória: vérnyomásmérő
Modell: 2004

Szüksége van segítségre?

Ha segítségre van szüksége Bioland 2004, tegyen fel kérdést alább, és más felhasználók válaszolnak Önnek




Útmutatók vérnyomásmérő Bioland

Bioland

Bioland 2006 Útmutató

16 Szeptember 2024
Bioland

Bioland 3006 Útmutató

19 Augusztus 2024
Bioland

Bioland 2004 Útmutató

16 Augusztus 2024
Bioland

Bioland 3005 Útmutató

13 Augusztus 2024
Bioland

Bioland 2005-2 Útmutató

14 Július 2024

Útmutatók vérnyomásmérő

Legújabb útmutatók vérnyomásmérő

Medisana

Medisana BPW 300 Útmutató

16 Október 2024
Omron

Omron RS1 Útmutató

4 Október 2024
Cresta

Cresta BPM701 Útmutató

4 Október 2024
Sanitas

Sanitas SBC 23 Útmutató

4 Október 2024
Beurer

Beurer PO 60 Útmutató

2 Október 2024
Medisana

Medisana MTP Pro Útmutató

2 Október 2024
Medel

Medel Palm Self Útmutató

28 Szeptember 2024
Medel

Medel Dualstetho PRO Útmutató

28 Szeptember 2024
Medel

Medel Stetho Self Útmutató

28 Szeptember 2024
Medel

Medel Sense Útmutató

27 Szeptember 2024